Categorii

AUTOMATIC FLASH AND FIRE POINT TESTER WITH BAROMETRIC CORRECTION OF THE RESULT

  • The automated Cleveland Open Cup flash point tester accurately determines flash and fire point temperatures of viscous petroleum products including oils and bitumens
  • The apparatus is a benchtop model which houses the mechanical components and a miniature PC with touch screen.
  • Flash and fire point detection through ionization sensor: the apparatus also provides an alert if a flash has occurred at the first flame application, warning that the test result is not reliable.
  • The flash point temperature remains shown on the display until the  operator’s acknowledgement.

The instrument is equipped with a sensor for barometric pressure for the correction of the results towards atmospheric pressure.


Producător: Scavini

Model: AD0092-600

Cod: 8aada460a716

Preţ, fără TVA: La cerere

Distribuie

Telefon: 0744.594.794 / 0251.417.664

E-mail: tehnic@aparatura-laboratoare.ro

  • The automated Cleveland Open Cup flash point tester accurately determines flash and fire point temperatures of viscous petroleum products including oils and bitumens
  • The apparatus is a benchtop model which houses the mechanical components and a miniature PC with touch screen.
  • Flash and fire point detection through ionization sensor: the apparatus also provides an alert if a flash has occurred at the first flame application, warning that the test result is not reliable.
  • The flash point temperature remains shown on the display until the  operator’s acknowledgement.

The instrument is equipped with a sensor for barometric pressure for the correction of the results towards atmospheric pressure.

Enamel finished steel case, benchtop version.

LAN connectivity, calibration and diagnostic

routines.

Electric heating plate.

 

Brass oil cup with insulating handle (n°1 Pt100 with cable and quick connector supplied with the instrument).

Safety device is provided to stop the analyzer ifa flash has not been detected at a temperature 30°C over the preset value or at a temperature of 400°C.

Electronically driven mechanism that passes the flame over the cup at the correct intervals for the correct amount of time

Dimensions (l x w x h): mm 360 x 460 x 680. Weight: kg 36 approx.

Second generation ionization flash sensor.

ASTM D 92 - IP 36 - ISO 2592

Flame extinguisher: automatically extinguishes the flame covering the

cup in case a fire is detected or after fire point detection.

CE marked

PC based controller with touch-screen interface.

Selectable cooling time, storage of up to 400 test results andpossibility to retrieve

and print test reports

Selection of the ASTM/IP test method or setup

of up to 40 custom methods

 

Possibility to change thesetpoint during the

test, “search” option

 

 

 

Denumire accesoriu

COD

Pret Euro fara TVA

Printer

AD0093-A00       

 

PT100 simulator

CAL001 

 

Officially certified PT100 simulator

CAL002 

 

CAPTCHA ImageRefresh Image